• Provides the electronics necessary to power and control SST’s range of zirconium dioxide (ZrO2 ) sensors
• Externally triggered automatic or manual calibration. Calibration can also be initiated via an on-board push button
• Power and sensor operating LEDs
• Removable polarised screw terminals for easy wiring
O2I-Flex Oxygen Sensor Interface Board BENEFITS
• Adaptive software filtering provides a fast sensor response coupled with a stable oxygen output
• High accuracy linear output
• Can be calibrated in fresh air (20.7% O2 ) or in any other known O2 concentration
O2I-Flex Oxygen Sensor Interface Board TECHNICAL SPECIFICATIONS
Supply voltage |
24V DC ± 10% |
Supply current |
600mA max. at 24V DC |
Digital output |
RS232 |
Analogue output |
4 ~ 20mA; load 100 ~ 600Ω or 0 ~ 10V DC ; load 10kΩ min |
Temperature limits Storage: Operating: |
-10°C to +70°C -10°C to +70°C |
Oxygen pressure limits |
1 ~ 1000mbar |
O2I-Flex Oxygen Sensor Interface Board OUTPUT VALUES
Oxygen range (analogue output) |
0.1 ~ 25% O2 or 0.1 ~ 100% O2 |
Oxygen range (RS232 output) |
0.1 ~ 100% O2 |
Accuracy after calibration |
1% O2 |
Repeatability after calibration |
0.5% O2 |
Output resolution: 0 - 10V DC: 4 - 20mA: RS232: |
0.01V 0.01mA 0.01% O2 |
Response time (step 10 - 90%) Fast response sensor connected: Standard response sensor connected: |
< 4s < 15s |
Initial warm up time (till stable output) |
5 ~ 10mins |
Output inactive start up delay (heater warm up) |
60s |