If you need to access a PID for cleaning or replacing, you must remove
the rubber boot and belt clip first. Please see FAQ for how to remove
the rubber boot for instructions on how to conduct that procedure. The
illustrated process for PID Sensor removal, cleaning and replacement can
be found in the Knowledge Base Manual/SOP entry- the steps are listed
below.
1. Turn off the instrument
2. Remove the four screws in holding the MultiRAE Pro sensor compartment cover in place
3. Remove the cover to expose the sensors
4. Gently lift out the PID module with your fingers
5. If the module requires replacement, place a new module into the slot,
being careful to match the indexing keys. The sensor can only go into
its slot one way
6. If you want to open the sensor module to inspect and clean the lamp
and sensor electode panel, you must use the special tool. Its “C” shaped
end has small teeth inside. Slide the tool so that the teeth slip into
the notch between the module’s cap and body
7. Gently pry up the cap using a rocking motion:
8. Once the cap is removed, set it aside
9. Now lift the sensor electrode panel from the module:
10. Clean the lamp’s window with a cleaning swab dipped in methanol lamp
cleaner, and allow it to dry. Do not touch the lamp window with your
fingers, as the residual oil will shorten its life.
11. Inspect the electrical contacts. Clean them with swab dipped in lamp cleaner if they appear to need cleaning
12. Reassemble the sensor module by placing the sensor electrode panel
back in place and firmly pressing the cap back onto the top
13. Place the sensor module back into the MultiRAE Pro. Make sure the index points are aligned.
14. Reinstall the rear cover
15. Tighten all four screws
ISweek(http://www.isweek.com/)- Industry sourcing & Wholesale industrial products