The objective of our study was to develop a novel capacitive force sensor that enables simultaneous measurements of yaw torque around the
pressure axis and normal force and shear forces at a single point for
the purpose of elucidating pressure ulcer pathogenesis and establishing
criteria for selection of cushions and mattresses.
Methods
Two newly developed sensors (approximately 10 mm×10 mm×5 mm (10) and 20
mm×20 mm×5 mm (20)) were constructed from silicone gel and four upper
and lower electrodes. The upper and lower electrodes had sixteen
combinations that had the function as capacitors of parallel plate type.
The full scale (FS) ranges of force/torque were defined as 0–1.5 N,
–0.5-0.5 N and −1.5-1.5 N mm (10) and 0–8.7 N, –2.9-2.9 N and −16.8-16.8
N mm (20) in normal force, shear forces and yaw torque, respectively.
The capacitances of sixteen capacitors were measured by an LCR meter
(AC1V, 100 kHz) when displacements corresponding to four degrees of
freedom (DOF) forces within FS ranges were applied to the sensor. The
measurement was repeated three times in each displacement condition (10
only). Force/torque were calculated by corrected capacitance and were
evaluated by comparison to theoretical values and standard normal force
measured by an universal tester.
Results
In measurements of capacitance, the coefficient of variation was 3.23%
(10). The Maximum FS errors of estimated force/torque were less than or
equal to 10.1 (10) and 16.4% (20), respectively. The standard normal
forces were approximately 1.5 (10) and 9.4 N (20) when pressure
displacements were 3 (10) and 2 mm (20), respectively. The estimated
normal forces were approximately 1.5 (10) and 8.6 N (10) in the same
condition.
Conclusions
In this study, we developed a new four DOF force sensor for measurement
of force/torque that occur between the skin and a mattress. In
measurement of capacitance, the repeatability was good and it was
confirmed that the sensor had characteristics that enabled the
correction by linear approximation for adjustment of gain and offset. In
estimation of forces/torque, we considered accuracy to be within an
acceptable range.
iSweek(http://www.isweek.com/)- Industry sourcing & Wholesale industrial products
没有评论:
发表评论