A re-calibration
method for a dual-beam NDIR gas sensor uses a calibration curve based upon a
combination of physics and sensor measurement components of the sensor to
calculate sample gas concentration, then determines a second gas concentration
measurement by a secondary gas standard which is used with a reversed
calibration curve algorithm to adjust the sensor measurement component. The
calibration curve is based upon a gamma ratio (“G”) that has been normalized by
G when no sample gas is present in the sample chamber (“G0”), G
being a ratio of a signal channel output (“VS”) of the sensor
divided by a reference channel output (“VR”) of the sensor. The
concentration (“P”) of sample gas in the sensor is calculated through use of
the calibration curve by a gas detection equation of P=F(x)=F(y/G0),
where x is a normalized ratio of VS/VR and y is G. The
reversed calibration curve algorithm is P=F(x)=F(y/G0N), where G0N=y1/x2,
y1=G for the sensor, x2=F−1 (P2)
and P2 is the second gas concentration of the sample gas.
Description
FIELD OF THE INVENTION
The present invention is in the field of measuring
instruments, and specifically relates to a method for re-calibrating
non-dispersive infrared (NDIR) gas sensors whose outputs have drifted over time
and no longer correctly reflect their measurement accuracy.
BACKGROUND OF THE INVENTION
Output stability or drift over time leading to
measurement inaccuracies has long been a major deficiency for gas sensors
irrespective of what technology or methodology is used for their conception or
realization. Output software correction may alleviate the problem somewhat but
it is in many instances inaccurate and not even always applicable. It has long
been the objective of many researchers in this field to overcome this problem
fundamentally and for good. Recently the present inventor in U.S.
application Ser. No. 12/859,749 advanced the teaching of an Absorption Biased
NDIR Gas Sensing Methodology which is capable of eliminating substantially all
NDIR gas sensor output drifts over time without the need for re-calibration
(Wong, filed 19-AUG-2010). As it turns out, the solution to solving this output
drift problem for gas sensors actually lies deeper than the availability of
superior NDIR gas sensor types even though they can indeed be designed to be
capable of maintaining measurement accuracy over time. The fact of the matter
is that people have experienced gas sensor output instability for such a long
time in the past that when output stable sensors really come along nobody would
believe it. Until such time that stable gas sensors become widely available and
users begin to consider their performance as trustworthy and truly believable,
the real need today must be viewed at a completely different perspective and
that is to be able to come up with a fast, inexpensive and simple methodology
that can easily check the accuracy of gas sensors and more importantly, just as
easy and simple, hence inexpensive, to re-calibrate them when they are found to
be inaccurate.
It is therefore the primary objective of the present
invention for the present author to advance a novel methodology to simply and
easily re-calibrate an NDIR gas sensor.
SUMMARY OF THE INVENTION
The present invention is generally directed to a
method for recalibrating a dual-beam NDIR gas sensor by using a calibration
curve that is based upon a combination of a physics measurement component of
the NDIR gas sensor and a sensor measurement component of the NDIR gas sensor
to calculate a first concentration of sample gas, then using a secondary gas
standard to determine a second gas concentration of the sample gas, and then
recalibrating the NDIR gas sensor by using the second gas concentration and a
reversed calibration curve algorithm which adjusts the sensor measurement
component to correct for any difference between the first concentration and the
second gas concentration when the difference between the two exceeds a
preselected threshold.
In a separate group of aspects of the present
invention, the calibration curve (which expresses the concentration of the
sample gas as an nth order, e.g., a third order polynomial of G) is based upon
a gamma ratio (“G”) that has been normalized by the gamma ratio when no sample
gas is present in the sample chamber (“G0”), G being a ratio of a
signal channel output (“VS”) of the NDIR gas sensor divided by a
reference channel output (“VR”) of the NDIR gas sensor. The
concentration (“P”) of sample gas in the NDIR gas sensor is calculated through
use of the calibration curve by a gas detection equation of P=F(x)=F(y/G0),
where x is a normalized ratio of VS/VR and y is G. The
reversed calibration curve algorithm, which is a non-linear equation, is
P=F(x)=F(y/G0N), where G0N=y1/x2, y1=G
for the NDIR gas sensor, x2=F−1 (P2) and P2
is the second gas concentration of the sample gas.
In another separate group of aspects of the present
invention, the secondary gas standard (which can be a second NDIR gas sensor)
is calibrated within a preselected time period prior to determining the second
gas concentration and both the first concentration and the second concentration
detect substantially the same concentration within a pre-selected space (e.g.,
a still space of less than 1,000 cubic feet). The second concentration can be
transmitted to the NDIR gas sensor being recalibrated which receives the
transmission and calculates gas concentration through electronics which use its
calibration curve and recalibrate the NDIR gas sensor and then provide
indication of recalibration.
Accordingly, it is a primary object of the present
invention to provide an improved methodology for re-calibrating NDIR gassensors whose outputs have drifted over time and no longer correctly reflect
their measurement accuracy.
This and further objects and advantages will be
apparent to those skilled in the art in connection with the drawings and the
detailed description of the invention set forth below.
没有评论:
发表评论