Two detector elements are optically isolated by having them mounted
(die-attached) on the same header so that the thermal tracking of the
detectors respectively for the signal and reference channels is close to
ideal. Furthermore, such an optical isolation technique or
cross-interference suppression between the two detector elements mounted
on the same header also allows the use of only one and the same narrow
band pass interference filter covering both detectors. Thus the thermal
tracking of the filters respectively for the signal and reference
channels is also close to perfection as both channels share the same
filter.
Description
CROSS-REFERENCE TO RELATED APPLICATIONS
The present application claims the priority benefit of U.S. Ser. No.
61/331,327, filed May 4, 2010, the disclosure of which is specifically
incorporated herein by reference. This application is also a
continuation-in-part of U.S. Ser. No. 12/859,749, filed Aug. 19, 2010,
the disclosure of which is specifically incorporated herein by
reference, which itself claimed the priority benefit of U.S. Ser. No.
61/274,874, filed Aug. 21, 2009, the disclosure of which is also
specifically incorporated herein by reference.
FIELD OF THE INVENTION
The present application is in the field of gas analysis, and
specifically relates to apparatus using a Non-Dispersive Infrared (NDIR)
gas analysis technique to determine the concentration of a particular
type of gas present in a chamber by sensing the absorption of infrared
radiation passing through the gas.
BACKGROUND OF THE INVENTION
The Non-Dispersive Infrared (“NDIR”) technique has long been considered
as one of the best methods for gas measurement. In addition to being
highly specific, NDIR gas analyzers are also very sensitive, stable,
reliable and easy to maintain and service. Ever since the NDIR technique
of gas measurement was first introduced and practiced in the mid
1950's, a large number of improved measurement techniques based upon the
NDIR principle for gas detection have been proposed and successfully
demonstrated. The most notable advances over the years in this field are
summarized as follows.
Burch et al. (U.S. Pat. No. 3,793,525) and Blau et al. (U.S. Pat. No.
3,811,776) in 1974 were the first to advance a so-called “Double Beam”
technique for NDIR gas measurement by taking advantage of the principle
of nonlinear absorption for some strongly absorbing gases such as CO2 to
create a reference channel. Shortly thereafter, this “Double Beam” NDIR gas sensor technique was greatly simplified with the use of two
interposed spectral filters (one absorbing and one neutral) to create a
sample and a reference detector channel. Subsequent NDIR gas sensors,
designed using this technique, have enjoyed good performance alluded to
briefly above.
In U.S. Pat. No. 4,578,762 (1986) Wong advanced the first
self-calibrating NDIR CO2 analyzer using a novel two-wheel chopper and
mirror arrangement. Another improved type of such gas analyzer is shown
and described in U.S. Pat. No. 4,694,173 (1987) by Wong. This gas sensor
has no moving parts for effecting the interposition of spectral filters
to create both a sample and reference detector channel as in the NDIR
gas analyzers described earlier.
In U.S. Pat. No. 5,163,332 (1992), Wong advanced the so-called
“wave-guide” sample chamber concept for simplifying NDIR gas sensors
into ones that are compact, rugged and low-cost while still maintaining
their superior performance characteristics. This concept has
subsequently been widely adopted in the design of today's NDIR gas
sensors, particularly in low-cost and high volume versions.
All of the NDIR gas analyzers described above for the measurement of the
concentrations of one or more gases in a mixture perform well
functionally and have contributed successfully to the overall technical
advancement in the field of gas analysis during the past two decades.
They have been widely accepted in both the medical and industrial
communities. Despite their undisputed success over the years, there
still remain a number of important sensor performance characteristics
that need to be greatly improved in order to further extend the useful
applications of these devices in a number of areas.
By far the most deficient performance characteristic of gas sensors of
today, inclusive of NDIR gas sensors, is the sensor output stability
over time. Unlike the temperature controller or thermostat device which
just about everybody is familiar with at home or in their workplaces for
sensing temperature that never requires output adjustment or
recalibration over time, such is not the case for gas sensors
irrespective of their operational principle, functional design, material
construct or even costs. Dependent upon the type of gas sensors, just
about every one of them requires recalibration once every six months to a
year without exception in order that they remain accurate over time.
While this performance deficiency has been well tolerated over the
years, it remains as a significant drawback for gas sensors and even
precludes their use in a number of vital applications and therefore
there has been a long-felt need for elimination of this problem.
The second most prominent performance deficiency for gas sensors of
today irrespective of their operational principle is their output
dependence as a function of the temperature of the environment wherein
the sensors are located. This performance deficiency for just about all
gas sensors is universally, albeit reluctantly, dealt with by specifying
the output correction per degree of temperature change with respect to
the output stipulated at a standard temperature. In some gas sensors
these output temperature corrections are quite large and in many cases
severely limit the use of these sensors outdoors. It would be a
significant step forward in the development of future gas sensors,
particularly for the NDIR type, because of its prevalent use in most
industries, that this performance deficiency be also overcome and,
again, there has been a long-felt need for overcoming this problem.
The afore-mentioned serious NDIR gas sensor performance deficiencies,
namely sensor output drift over time and output dependency as a function
of exposed sensor temperature, have earlier been addressed by the
present inventor in a provisional patent application 61/274,874 to the
US Patent Office filed on Aug. 21, 2009 and entitled “Absorption Biased
NDIR Gas Sensing Methodology.” In this recent patent disclosure, the
present inventor takes advantage of the fact that if the spectral
content of radiation from the source and/or convoluted with those from
the surroundings be always kept the same for both the reference and the
signal channels of an NDIR gas sensor, assuming that this sensor uses
the most widely deployed dual-channel methodology, the output of the
sensor taken as the ratio of the signal output over the reference output
can always be kept constant or unchanged over time except when the gas
of interest is present in the sample chamber.
In order that this recently disclosed Absorption Biased methodology be
implemented, both the signal and the reference channel must be provided
with exactly the same spectral narrow band pass filter designed for
detecting the gas of interest in front of the respective infrared
detectors. In order to differentiate between the signal and the
reference channel outputs from the respective detectors in the presence
of the gas of interest, an absorption bias is designed between the two
channels via the use of different sample chamber path lengths for the
two channels. Thus, if the sample chamber path length for the signal
channel is longer than that for the reference channel, the signal
channel detector output will change greater (or be reduced more) than
that for the reference channel when the same concentration level of the
gas of interest is present in the sample chamber. In other words, the
sensor output will change as the concentration level of the gas of
interest changes in the sample chamber as reflected by the calibration
curve which can be prepared for the sensor.
The fact that both detection channels have the same narrow band pass
spectral filter and they receive radiation from one and the same single
infrared source as taught by the widely deployed dual-channel NDIR gas
detection methodology, they are all affected in the same way to first
order when there are spectral changes caused by temperature variations
in the sample chamber and/or by the short or long-term operational
changes (e.g. aging) of the infrared source. Thus the outputs of the
dual-channel NDIR gas sensor for the detection of any gas of interest
implemented using the inventor's recently disclosed Absorption Biased
methodology will stay virtually drift-free over time without the need
for any periodic re-calibration or software correction.
ISweek(http://www.isweek.com/)- Industry sourcing & Wholesale industrial products
没有评论:
发表评论