A gas sample separated from transformer oil is circulated through an
NDIR gas sensor system which obtains an acetylene concentration by
calculating a detected acetylene concentration obtained by an absorption
biased (“AB”) NDIR acetylene gas sensor, calculating a detected carbon
dioxide concentration obtained by an AB NDIR carbon dioxide gas sensor,
calculating a detected water vapor concentration obtained by an AB NDIR
water vapor NDIR gas sensor and then determining the acetylene
concentration from the detected acetylene concentration through use of
the detected carbon dioxide and water vapor concentrations to compensate
for their interference.
DETAILED DESCRIPTION OF THE INVENTION
The present invention advances a methodology for the design of an NDIR
gas sensor system which is capable of detecting sub-ppm levels of
Acetylene gas in transformer oil in the presence of multiple fault gases
and moisture. In order to achieve a successful methodology, a number of
technical obstacles must be recognized, addressed and overcome.
The present invention addresses the drawback of NDIR gas sensors for
detecting gases in the sub-ppm concentration ranges having very weak
absorption bands located in the far infrared region of the
electromagnetic spectrum (>12.0μ) and very unique ultra-narrow line
shapes and overcomes these drawbacks by a novel NDIR design methodology.
This design methodology comprises 1) matching the ultra-narrow line
shape of the gas' absorption band with a carefully designed narrow band
pass filter; 2) making sure that the out-of-band leakage of the filter
is not more than one part in 103; 3) utilizing a MEMS infrared source
with built-in collimating optics and a large area thermopile detector
for S/N optimization; 4) using a sufficiently long path length waveguide
sample chamber for achieving adequate modulation by the target gas and
5) including a sensitive CO2 and a sensitive Dew Point NDIR sensor for
simultaneously measuring the concentration levels of both gases in order
to compensate for any possible interferences by these gases to the
measurement of the target gas.
The first step needed to advance a successful methodology is to design a
narrow infrared band pass filter with a center wavelength (CWL)
specified at 13.70μ coincident with the center wavelength of the
absorption line of Acetylene and with an adequately narrow full width at
half-maximum (FWHM). The FWHM of the specified narrow band pass filter
must be narrow enough so as to afford adequate signal modulation for
measuring concentration levels down to sub-ppm or ppb for Acetylene gas.
In order to be able to detect Acetylene down to +/−0.5 ppm
concentration level, a thermopile detector is used with an average D*
around 1×108 cm Hz0.5 W−1 and the FWHM of the narrow band pass filter
must not exceed 0.1μ.
With the required spectral properties for the narrow band pass filter
determined, step two is to calculate the path length for the sample
chamber required to be able to accurately measure concentration levels
of Acetylene down to +/−0.5 ppm. Calculations showed that based upon the
very weak absorption of Acetylene gas, the minimum path length required
to get the job done is no less than ˜20 inches.
The third step is to recognize the possibility of interference effects
on the measurement accuracy for Acetylene by other gases and then
compensate for any such interference. After a careful study it was found
that only two gases, namely carbon dioxide (CO2) and water vapor (H2O),
could severely interfere with the measurement accuracy of Acetylene at
13.70μ. For CO2, it is due to the presence of a very strong absorption
band at ˜15.0μ. Water vapor on the other hand has absorption bands
almost everywhere in the infrared extending way into the far infrared
wavelength region. Even though water vapor has several very strong
absorption bands in the infrared, those that are present around 13.70μ
are extremely weak. Due to the fact that the FWHM of the specified
filter is only ˜0.1μ, it is not expected that water vapor will cause any
interference problem. However, the state-of-the-art for making a narrow
band pass interference filter in the far infrared today only guarantees
spectral blocking outside of the designed pass band to be no better
than 1:103. Thus, because of the omnipresence of water vapor absorption
lines everywhere, the out-of-the-band absorption, i.e. for the
wavelength region outside of the FWHM at 13.70μ, the interference effect
of water vapor is still very significant. Furthermore, because of the
fact that the amount of water vapor present in air or dissolved in the
transformer oil could vary from a few to tens of mmHg (1 mmHg=1,316 ppm
at Standard Temperature and Pressure [STP] conditions), its interference
effects on the measurement accuracy of Acetylene cannot be ignored. In
conclusion, after a careful analysis of the subject, the interference
effects of both CO2 and water vapor must be carefully taken into
consideration before the measurement accuracy of Acetylene down to
sub-ppm levels can be confidently realized.
To take the interference effects of both CO2 and water vapor on the
measurement accuracy of Acetylene into consideration is in theory not a
complicated task. One practical and viable approach is to first
ascertain quantitatively, based upon the narrow band pass filter to be
used in the design of the Acetylene sensor, the amount of Acetylene that
is equivalent to a certain known amounts of both CO2 and water vapor.
The experimentally measured results indicated that under the same
measurement conditions, the presence of 5,000 ppm of CO2 would be
equivalent to the presence of 20 ppm of Acetylene and the presence of
8.0 mmHg of water vapor (H2O) would be equivalent to 40 ppm of
Acetylene. Thus, in order to be able to guarantee the accuracy of +/−0.5
ppm of Acetylene, one must be able to accurately detect +/−5,000/40 or
+/−125 ppm of CO2 and +/−8.0/80 mmHg or +/−0.1 mmHg of water vapor
simultaneously while measuring Acetylene using the same sensor in order
to adequately compensate for their presence as interfering gases.
As it turns out, it is possible to design an output stable NDIR CO2 gas
sensor capable of detecting the gas with an accuracy of +/−100 ppm and
with a reasonable response time (0-90%) commensurate with that required
for measuring Acetylene which is 3-5 minutes. The challenge lies in the
fact that an output stable NDIR dew point or water vapor sensor capable
of detecting the gas with an accuracy of +/−0.1 mmHg simply cannot be
found anywhere today. Using a conventional humidity sensor and
converting its output to water vapor pressure by measuring also the
temperature will not work because of two factors. First, humidity
sensors are known to have output drifts that are difficult to determine
over time. Second, presently available humidity sensors are not accurate
enough to meet the required accuracy of +/−0.1 mmHg. Thus, the present
invention has to embark on an additional innovative step to realize an
NDIR gas sensor capable of measuring water vapor pressure with an
accuracy of +/−0.1 mmHg and a response time (0-90%) of 3-5 minutes.
Finally in order to complete a well-thought-out measurement methodology,
the best available infrared source and thermopile detector must be
selected for the design of the sensor. At the same time the accompanying
signal processing electronic system should also be designed to be
detector-noise-limited. In other words, the overall noise of the sensor
can only be limited by the detector and not by the infrared source, the
electronic signal processing circuit or any part of the remaining
system.
ISweek(http://www.isweek.com/)- Industry sourcing & Wholesale industrial products
没有评论:
发表评论